To Adjust Scale And Offset With The Touchscreen; To Specify Channel Coupling - Keysight Technologies IntegraVision PA2200 Series Operating And Service Manual

Power analyzers
Table of Contents

Advertisement

To adjust scale and offset with the touchscreen

You can also adjust the vertical scale and offset using the touchscreen. For example, below, the
user has touched the 2.00 A/ indicator at upper left to open the Channel 1 sidebar on the right.
You can use this sidebar to configure the channel's scale, offset, and coupling.

To specify channel coupling

You can specify each channel's input coupling as AC or DC. If the channel is DC coupled, you
can quickly measure the DC component of the signal by simply noting its distance from the
ground symbol. If the channel is AC coupled, the DC component of the signal is removed,
allowing you to use greater sensitivity to display the AC component of the signal.
To configure coupling, press [1] or [2] (or [3] or [4] on the PA2203A), and then use the Coupling
softkey to make your selection.
48
Keysight IntegraVision PA2200 Series Power Analyzers Operating and Service Guide