Updating Firmware

From time to time, Keysight Technologies may release firmware updates for the instrument. To
see the instrument's currently installed firmware version, press [Help] > About IntegraVision .
To perform the firmware update:
1. Put the USB flash drive with the *.KUP file into one of the front-panel USB ports.
2. Press [Utility] > File Explorer .
3. Use the first softkey and the Entry knob to browse to the *.KUP file on the USB drive.
4. Press the Load File softkey.
The firmware update utility will run, and the instrument will cycle power to start running the new
firmware.
176
Before updating instrument firmware, remove power from the power channel
inputs. You may leave the connectors in place, but there should be no current
or voltage signals going into the power analyzer.