Making Selections In Setup Mode - Nonin 9847V Operator's Manual

Veterinary handheld pulse; oximeter and carbon dioxide (co2) detector
Table of Contents

Advertisement

Using the 9847V Pulse Oximeter
Entering Setup Mode
1. With the unit in Standby mode, press and hold Alarm Limit Review while pressing and then
releasing On/Standby.
2. Release Alarm Limit Review when 888 888 displays on the SpO
displays. Three brief beeps sounds, and ALr and dFt appear in the SpO
displays.
NOTE: Entering setup mode with audible alarms disabled:
1. Press and hold Alarm Limit Review and Audible Alarm Disable while pressing and
releasing On/Standby.
2. Release Alarm Limit Review and Audible Alarm Disable. The display appears as
described in #2 above, and the Audible Alarm Limit indicator is steadily lit.
NOTE: You may exit setup mode at any point and save your current animal alarm limit changes
(without stepping through the remainder of the setup mode menu options).
1. Complete the desired selections, then turn off the 9847V.
2. Next, enter setup mode again and choose rCL (recall stored limits) from the alarm mode
settings.
NOTE: Setting the month to "00" disables the calendar and clock functions and helps conserve
battery life.
NOTE: Setup mode is disabled if the batteries are critically low at power on.

Making Selections in Setup Mode

1. Upon entering setup mode, "ALr" (the first parameter) appears in the SpO
Advance (or press and hold to quickly scroll) to increment the range of values on the pulse
rate display. The menu starts at the current value stored in memory for the parameter
designated in the SpO
2. When the value appears in the pulse rate display, press Alarm Limit Review to store the
value and advance the SpO
alarm limit, etc.) as listed in Table 2.
3. Continue this process until all parameters are set.
4. When the setting sequence has been completed, the 9847V:
a. Exits setup mode.
b. Briefly displays rCL ALr (recall alarm settings), followed by the alarm limit settings.
c. Begins normal operation.
display and cycles through the range of values listed in Table 2.
2
display to the next sequential parameter (no breath delay time,
2
and pulse rate
2
and pulse rate
2
display. Press
2
19