No Breath Delay Time Setting; Spo Low Alarm Limit Setting; Pulse Rate Alarm Limit Setting; Spo 2 High Alarm Limit Setting - Nonin 9847V Operator's Manual

Veterinary handheld pulse; oximeter and carbon dioxide (co2) detector
Table of Contents

Advertisement

No Breath Delay Time Setting

1. Upon entering no breath delay time setup, "nbd" appears in the SpO
delay time may be set from 15 to 60 (seconds) in increments of 5, or to OFF.
2. When the no breath delay time has been selected, setup mode continues to the SpO
alarm limit setting (C02 alarm mode goes to FPt next, then to y, etc.).
SpO
Low Alarm Limit Setting
2
1. Upon entering the SpO
alarm limit may be set from 50 to 95 (%) in increments of 1, or to OFF.
2. When the SpO
low alarm limit setting has been selected, setup mode continues to the pulse
2
rate alarm limit settings.

Pulse Rate Alarm Limit Setting

1. Upon entering the pulse rate low alarm limit, "H L" appears in the SpO
rate low alarm limit may be set from 20 to 200 (beats per minute) in increments of 5, or to
OFF.
2. After selecting the pulse rate low alarm limit, the SpO
setup mode for the pulse rate high alarm limit. The pulse rate high alarm limit may be set to
one of the following:
• 100 to 425 (beats per minute) in increments of 25
• OFF
• 50 to 100 (beats per minute) in increments of 5
3. When the desired pulse rate high alarm limit setting has been selected, setup mode
continues to the SpO
SpO
High Alarm Limit Setting
2
1. Upon entering the SpO
alarm limit may be set from 80 to 100 (%) in increments of 1, or to OFF.
2. When the SpO
high alarm limit setting has been selected, setup mode continues to the
2
breath beep pitch settings.

Breath Beep Pitch Setting

1. Upon entering the breath beep pitch, "FPt" appears in the SpO
pitch defaults to variable pitch (fixed pitch off). Refer to "Breath Beep" on page 27 for
additional information.
2. The breath beep pitch selection may be toggled between 0N or 0FF.
22
low alarm limit, "02L" appears in the SpO
2
high alarm limit settings.
2
high alarm limit, "02H" appears in the SpO
2
Using the 9847V Pulse Oximeter
display. The no breath
2
display. The SpO
2
display. The pulse
2
display shows "H H" indicating the
2
display. The SpO
2
display. The breath beep
2
low
2
low
2
high
2

Advertisement

Table of Contents
loading