Table 4: Audible and Visible Indicator Functions During Alarm Conditions (Continued)
Alarm Condition
SpO
high or low
2
Pulse rate high or low
Inadequate pulse quality
No breath (no breath
detected during set delay
time interval measured in
seconds)
Pulse oximeter sensor
disconnect, dislodgment,
or pulse oximeter sensor
failure
CO
sensor reduced
2
signal due to:
• CO
sensor
2
disconnection
• Malfunction
• Blocked light path
• Airway adapter tube
removed from the CO
sensor
• Airway adapter tube
improperly installed in
airway sensor
a. Note that if a critical battery condition occurs while the audible alarms are disabled, the audible alarms will become re-
enabled. A medium priority alarm will sound unless there is a high priority alarm condition in effect.
30
Audible Indication
High priority animal alarm SpO
High priority animal alarm Pulse rate numeric display blinking; if latched
High priority animal alarm Pulse Quality indicator blinks red as it
High priority animal alarm • No Breath indicator flashing.
Medium priority alarm
until valid signal again
detected, unless audible
high priority animal alarm
is latched.
Medium priority alarm
until valid signal again
detected, unless audible
high priority animal alarm
is latched.
2
Using the 9847V Pulse Oximeter
Visible Indication
numeric display blinking; if latched for
2
critical battery state, displays blinking dashes
in all 3 LEDs.
for critical battery state, displays blinking
dashes in all 3 LEDs.
normally does in inadequate pulse quality
situations; can be latched for critical battery
state.
• CO
bar graph remains active.
2
• CO
bar graph blank (disabled) during
2
critical battery state.
• Display "dash" sign in leftmost SpO
and freeze numeric displays for 10
seconds.
• Display dashes in middle digit of SpO
and pulse rate displays 10 seconds after
first dash.
• Corresponding dashes will blink if prior
high priority pulse oximeter animal
alarms are latched.
• Third and sixth bars of CO
steadily lit.
• Latched prior No Breath indicator blinks.
digit
2
2
display
2