Airway Adapter Tube

WARNING: Do not reuse the Model 9840AAT Airway Adapter Tube. Cleaning the
interior will damage the anti-fog coating and cause inaccurate readings or a CO
sensor alarm.
WARNING: The Model 9840AAT Airway Adapter Tube will increase dead space by
approximately 6 cubic centimeters (0.4 cubic inches); this may adversely affect
ventilation for animals with small tidal volumes.
CAUTION: Do not use the airway adapter tube if the airway adapter tube is below
!
5 °C (41 °F). An airway adapter tube that is below 5 °C (41 °F) may frost, causing a
false reading. Warm the airway adapter tube to above 5 °C (41 °F) by putting it in a
warm place (for example in your hands or in a vehicle) before use.
CAUTION: An airway adapter tube that is between 5 °C (41 °F) and 10 °C (50 °F)
!
may cause inaccurate readings due to fogging of optical surfaces. It is recommended
that the airway adapter tube be warmed to above 10 °C (50 °F) before use.
The Model 9840AAT Airway Adapter Tube (figure 8) is a single-use only, disposable adapter designed
to be placed between the endotracheal tube and the breathing circuit of intubated animals. The airway
adapter tube connects to the CO
The CO
detector will not function properly unless the light emitting and detecting elements of the CO
2
sensor are properly aligned with the windows in the airway adapter tube.
(To Breathing
Device)
Attaching the Airway Adapter Tube to the CO
1. While grasping the large end of the airway adapter tube, place the clear windows of the tube
toward the CO
sensor (figure 9-A). The reflector should face away from the CO
2
NOTE: It is possible to force the airway adapter tube and the CO
alignment and connection. However, the CO
pieces are correctly attached to each other.
34
sensor.
2
Windows
Figure 8: Model 9840AAT Airway Adapter Tube
detector will not function properly unless these
2
Carbon Dioxide Sensor and Airway Adapter Tube
(To Endotracheal
Tube)
Tabs
Reflector
Sensor
2
sensor into an improper
2
2
2
sensor.
2