Power Requirements

The only physical installation of your Agilent spectrum analyzer is a connection to
a power source.
Line voltage does not need to be selected.
This analyzer does not contain customer serviceable fuses.
WARNING
Failure to ground the analyzer properly can result in personal injury. Before
turning on the analyzer, you must connect its protective earth terminals to the
protective conductor of the main power cable. Insert the main power cable
plug into a socket outlet that has a protective earth contact only. DO NOT
defeat the earth-grounding protection by using an extension cable, power
cable, or autotransformer without a protective ground conductor.
If you are using an autotransformer, make sure its common terminal is
connected to the protective earth contact of the power source outlet socket.
This is a Safety Class 1 Product (provided with a protective earthing ground
incorporated in the power cord). The mains plug shall only be inserted in a
socket outlet provided with a protective earth contact. Any interruption of the
protective conductor inside or outside of the product is likely to make the
product dangerous. Intentional interruption is prohibited.
To prevent electrical shock, disconnect the Agilent spectrum analyzer from
WARNING
mains before cleaning. Use a dry cloth or one slightly dampened with water to
clean the external case parts. Do not attempt to clean internally.
CAUTION
VENTILATION REQUIREMENTS: When installing the product in a cabinet, the
convection into and out of the product must not be restricted. The ambient
temperature (outside the cabinet) must be less than the maximum operating
temperature of the product by 4C for every 100 watts dissipated in the cabinet. If
the total power dissipated in the cabinet is greater than 800 watts, then forced
convection must be used.
This analyzer has autoranging line voltage input. Be sure the supply voltage is
within the specified range.
NOTE
For detailed analyzer specifications, see the Specifications guide.
Chapter 1
Installation and Setup

Power Requirements

11