Electronics Verification
March 2016
C.3.2

SI units

Example 1
In this application, an input frequency of 80 Hz represents a flow rate of 1,686.8 lpm.
Given
Other conditions
Equation
Calculations
Example 2
In this application, an input frequency of 650 Hz represents a flow rate of 3,446.4 kg/hr.
Given
Electronics Verification
www.aotewell.com
Q = flow rate
F = frequency: 80 Hz
K = K-factor (compensated): 10.772 (via fieldbus configuration tool)
C
= C
x
8
C
= 4.40287E-03 (from
8
Fluid = water
Line size = 80 mm
Line pressure = 700 kPag
Operating temperature = 60 °C
Q
=
------------------------- -
=
Q
10.772
-------------------------------------------
Q
=
×
10.772
=
Q
M = mass flow rate
F = frequency: 650.00 Hz
K = K-factor (compensated): 10.715 (via fieldbus configuration tool)
C
= C
x
16
C
= C
/ρ (from
16
11
ρ = density: 4.169 lb/m
3
sales@aotewell.com
Table C-1 on page
F
-------------- -
×
K
C x
80
×
C 8
80
0.00440287
1,686.8 lpm
Table C-1 on page
156)