Electromagnetic Immunity - Philips EPIQ 7 User Manual

Ultrasound system
Table of Contents

Advertisement

Safety
When connecting other accessories to the system, such as a remote video monitor or
computer, it is the user's responsibility to ensure the electromagnetic compatibility of the
system. Use only CISPR 11 or CISPR 22, Class B-compliant devices, unless otherwise noted.
WARNING
Using cables, transducers, and accessories other than those specified for use with the system
may result in increased emissions from, or decreased immunity of, the system.
CAUTION
Use only Class-B USB storage devices with the system. Some plastic-cased unshielded USB
devices may cause RF emissions that exceed Class-B limits. See the device's documentation to
determine whether it is Class-B compliant.
Approved Accessories
Accessory
Ultrasonic imaging transducer
DVD recorder
Black-and-white printer
Color printer

Electromagnetic Immunity

The system is intended for use in the electromagnetic environment specified here. The
customer or the user of the system should ensure that it is used in such an environment.
NOTE
The guidelines specified here may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
74
Manufacturer
Philips
Sony
Sony
Sony
Electromagnetic Compatibility
Model Number
Use only the transducers listed in
"Clinical Applications and
Transducers" on page 174
DVO-1000MD
UP-D897MD
UP-D25MD
EPIQ 7 User Manual 4535 617 25341