Biopsy Guideline Display - Philips EPIQ 7 User Manual

Ultrasound system
Table of Contents

Advertisement

Biopsy Guides

Biopsy Guideline Display

WARNING
Inspect all components and the transducer. Ensure that the biopsy guide you are using is the
correct one for the transducer, the system, and system software. Your Philips representative
can verify this information for you.
WARNING
Use only Philips-approved biopsy guides, transducers, covers, brackets, supplies,
components, and accessories. Other brands may not properly fit Philips transducers.
Improper installation may result in patient injury.
WARNING
Some biopsy guides must be installed over a sterile transducer cover.
WARNING
After each use, biopsy guides must be either sterilized or disposed of, depending upon the
type. See the instructions included with the biopsy guide.
WARNING
Most transducers can only be disinfected; they cannot be sterilized. Only the transducer
cover provides the sterile barrier.
Biopsy Guideline Display
WARNING
Do not attempt to use the biopsy guide until you have read the instructions for selecting the
display, installing the sterile transducer cover, and verifying alignment of the biopsy guide.
246
EPIQ 7 User Manual 4535 617 25341