Using the X7-2t TEE Transducer
Component

X7-2t Deflection Controls

The smaller knob controls medial/lateral movement, while the larger knob controls anterior/
posterior movement. To place the tip of the TEE transducer into the neutral position, align the
ribs on each knob with the center of the array rotation buttons (as shown below).
The knobs can be controlled by a detent brake that holds the tip position without locking it in
place. This allows the tip to straighten if it meets additional resistance. When the detent brake
actuator is rotated to the right (as shown) both knobs are in the freewheeling mode. When the
detent brake actuator is centered, the small knob (medial/lateral movement) is in the detent
mode, and when the actuator is rotated to the left, both knobs are in the detent mode.
EPIQ 7 User Manual 4535 617 25341
Transesophageal Transducers
Description
Transducer handle
Deflection controls