Preparing the System
Connecting Devices
WARNING
Off-cart devices connecting to the ultrasound system must comply with the applicable IEC or
national standards, such as IEC 60601-1, IEC 60950, or the equivalent.
CAUTION
Using accessories, transducers, peripherals, or cables not supplied with the ultrasound
system or recommended by Philips can affect the system in the form of increased emissions
or decreased immunity to external EMI/EMC occurrences. Non-specified peripherals, and
cables in some cases, can also increase leakage current or compromise the safety of the
grounding scheme.
CAUTION
If systems, transducers, and peripherals have been in an environment below 10°C (50°F),
allow them to reach room temperature before connecting or turning them on. Philips
recommends allowing 24 hours for complete normalization. Otherwise, condensation inside
the devices could cause damage. If the device was only briefly exposed to temperatures
below 10°C (50°F), then the time required for the device to return to room temperature could
be significantly less than 24 hours.
NOTE
Any device that is not purchased from Philips or a Philips-authorized agent is not covered under
a Philips service agreement or warranty.

External Printers

You can connect different external printers to your system.
WARNING
Images printed on a report printer are intended only for reference and should not be used for
diagnostic purposes.