1. GENERAL INFORMATION

1.1 Danger, precautions and warranty

Before putting the centrifuge into operation, please read this instruction manual
carefully.
This device may only be operated by trained specialist stuff. They must have carefully read
the operating manual and be familiar with the function of the device.
To protect people and environment the following precautions should be observed:
During centrifugation, the presence of people and the setting up of hazardous materials is
prohibited within 30 cm around the centrifuge according to the regulations of
EN 61010-2-020.
The HERMLE SIEVA 3 is non explosion-proof and must therefore not be operated in
explosion-endangered areas or locations. Centrifugation of flammable, explosive, radioactive,
or such substances, which chemically react with high energy, is strictly prohibited. The final
decision on the risks associated with the use of such substances is the responsibility of the
user of the centrifuge.
Never spin toxic or pathogenic material without adequate safety precautions, i.e.
centrifugation of buckets / tubes without or with defective hermetic sealings is strictly
prohibited. The user is obliged to perform appropriate disinfection procedures in case
dangerous substances have contaminated the centrifuge and or its accessories. When
centrifuging infectious substances, always pay attention to the General Laboratory
Precautions. If necessary, contact your safety officer!
It is prohibited to run the centrifuge with rotors other than listed for this unit.
Under no circumstances open the lid of the centrifuge while the rotor is still running or rotating
with a speed of > 2m/s
The following rules must be strictly adhered to:
Do not operate the centrifuge in case it is not installed correctly.
Do not operate the centrifuge when dismounted (e.g. without housing).
Do not run the centrifuge when mechanical or electrical assembly groups have been
tampered with unauthorized persons.
Do not use accessories such as rotors and buckets, which are not exclusively approved by
HERMLE Labortechnik GmbH, except commercially available centrifuge tubes made of glass
or plastic.
Do not spin extremely corrosive substances, as they may cause material damages and impair
mechanical resistance.
Do not operate the centrifuge with rotors or buckets, which show any signs of corrosion or
mechanical damage.
The manufacturer is responsible for safety and reliability of the centrifuge, only if:
The unit is operated in accordance with this instruction manual.
Modifications, repairs or other adjustments are performed by HERMLE-authorized personnel
and the electrical installation of the related location corresponds to the IEC-regulations.
© Hermle Labortechnik GmbH
Sieva-3_V1.18_eng