Position The Table Longitudinally And Transversely - GE Optima XR646 Operator's Manual

Digital radiographic system
Table of Contents

Advertisement

3. Hold the foot pedal down until the desired height is reached.
Note: The tabletop automatically stops when it reaches its maximum height of 825 mm (32.0 inches).
4. Remove your foot from the pedal to stop the movement.
5. To lower the table, press the Down pedal two consecutive times ("double-tap").
Figure 8-45 Down pedal
6. Hold the foot pedal down until the desired height is reached.
Note: The tabletop automatically stops at a safety point to prevent toe pinches. To reach the minimum
height, double-tap the foot pedal again and the system will track to the lowest position.
7. Remove your foot from the pedal to stop the movement.

Position the Table Longitudinally and Transversely

4 way table position
The standard table position can be adjusted in the longitudinal and transverse directions for greater
patient positioning flexibility. Use this procedure to position the tabletop in the longitudinal and trans-
verse directions with respect to the X-ray tube.
1. Release the table lock, if necessary.
2. Press the table top positioning pedal
the foot pedal.
Figure 8-46 Table top positioning foot pedal
3. Hold the foot pedal down and position the table top.
 You can float the table top in all directions while the pedal remains held down.
4. Manually move the tabletop in a longitudinal or transverse direction to the desired position.
5. Release the foot pedal to lock the tabletop.