GE Optima XR646 Operator's Manual page 46

Digital radiographic system
Table of Contents

Advertisement

CAUTION
Refer to the table below for maximum attenuation equivalent of possible materials located in
the x-ray beam.
Table 2-7 Maximum Attenuation Equivalent mm AL
Image Pasting patient barrier
Non-elevating table
Non-elevating table with carbon-fiber table top
Digital elevating table with floating table top
Weight-Bearing Stand
Standard elevating table with floating table top
Manual Wallstand front panel
Standard Arm Wallstand front panel
Extended Arm Wallstand front panel
CAUTION
5495975-1EN Rev.9
© 2013-2017 General Electric Company. All rights reserved.
Attention to the possible adverse effect arising from materials located
in the X-ray beam.
Item
Residual radiation hazard! Do not stand behind image receptor.
Chapter 2: Safety and Regulatory
mm Al
<0.08
<1.0
<1.0
<0.7
1.08
<1.0
<0.7
<0.7
<0.7
2-19