Standard Table
The standard table includes the digital receptor, or Digital Detector removable grid, foot pedals, and
Emergency Stop buttons.
Figure 8-43 Standard Table Components
Components
Table 8-14 Standard Table Components
Item
1. Tabletop
2. Elevating Base with
Telescopic Covers
5495975-1EN Rev.9
© 2013-2017 General Electric Company. All rights reserved.
The table dimensions are 2330mm (92 in) in length and 838mm (33 in) in
width. Its inherent filtration is less than 1.0 mm of aluminum equivalent at
100 kVp. The tabletop can be moved longitudinally, transversely and
elevated for easy patient positioning and can support a patient weighing
up to 250 kg (552lbs).
Tabletop is identified as applied part.
The elevating base raises and lowers the tabletop from 525mm (+10mm)
to 825mm (-10mm) from the floor level (21 in - 32 in).
The telescopic covers are assembled in two levels. Their purpose is to
cover the table power supply and the electronic and mechanical
components located in the table base. This is essential when the tabletop
is raised or lowered.
Chapter 8: System Hardware Overview
Description
8-59