GE Optima XR646 Operator's Manual page 88

Digital radiographic system
Table of Contents

Advertisement

Chapter 4: General Information
Figure 4-6 Finger pinch lock trigger
The system will beep and a message will appear on the Acquisition Workstation informing you of the
table lock. Exposures will be inhibited until the lock is resolved.
To unlock the table top:
1. Release the Table top positioning foot pedal (if necessary).
Figure 4-7 Table top positioning foot pedal
2. Remove hands, fingers, or other object from under the table edge.
3. Press the table top positioning foot pedal
(Figure
4-7) two consecutive times ("double-tap").
4. Hold the foot pedal down and position the table top.
5495975-1EN Rev.9
4-8
© 2013-2017 General Electric Company. All rights reserved.