Table 2-18 Standard Table warning Labels Icons
Item
Description
1
Table pinch point label
2
Patient load label
WARNING
1. The table can be moved vertically when the load is less than or equal to 180kg (397lbs),
located in the center of the tabletop end and the tabletop is positioned in the center.
2. The table can be moved vertically when the distributed load over the table is less than or
equal to 250kg (552lbs) and the tabletop is positioned in the center.
CAUTION
When the patient weight exceeds 180kg (397lbs) then patient on/off load of
the table should only occur from the center of the front or rear side of the
table (Maximum weight 250kg (552lbs)).
5495975-1EN Rev.9
© 2013-2017 General Electric Company. All rights reserved.