Rotate The Multi-Leaf Collimator; General Acquisition - GE Optima XR646 Operator's Manual

Digital radiographic system
Table of Contents

Advertisement

CAUTION
Potential Pinch Point:
a pinch point when the tube is rotated. Operators should keep their hands on the
OTS handle and keep patient's clear while rotating the tube.

Rotate the Multi-Leaf Collimator

1. Move the locking lever on the multi-leaf collimator toward the front panel, i.e. toward you. This
releases the collimator from the 0 lock-in position.
2. Grasp the multi-leaf collimator with both hands.
3. Rotate it to the desired angle and direction.
4. Return the locking lever to its original position. This prevents further rotation.
Note: When the collimator is rotated, the image border is adjusted to the maximum size based on the
selected collimator FOV.
CAUTION
Always grasp the multi-leaf collimator in such a way that neither hand can be
pinched or crushed between the handles and the collimator.

General Acquisition

This section outlines the basic acquisition process.
Refer to
Chapter 9: Worklist
form.
Refer to
Chapter 10: Image Acquisition
tions, how to conduct specific types of exams, and how to end exams.
1. Add or select the procedure from the Worklist.
CAUTION
Select all procedures for the patient.
2. Click [START EXAM] or [RESUME EXAM].
3. Select the protocols for the exam.
4. Click [ACCEPT].
5. If multiple procedures were selected from the Worklist, choose the exam to perform.
6. Choose the protocol to perform.
5495975-1EN Rev.9
© 2013-2017 General Electric Company. All rights reserved.
The area where the tube connects to the column may create
for more detailed information about adding or selecting procedures to per-
for more detailed information about the Acquisition screen func-
Chapter 5: Quick Steps