Certificates and approvals
SITRANS P500 with HART
Operating Instructions, 12/2014, A5E02344528-09
Connection
Type of protection "n" (Zone 2)
Designation
Connection "nA, ic"
Connection "nL"
Effective inner inductance
Effective inner capacitance
Explosion protection for USA
(to FM)
Certificate of Compliance
Marking (XP/DIP) or (IS)
Permitted ambient temperature T
Entity parameters
Marking (NI/S)
Permitted ambient temperature T
NI/S parameters
Explosion protection for Cana-
da (to
CSA
)
C
US
Certificate of Compliance
Marking (XP/DIP)
Permitted ambient temperature T
Entity parameters (XP/DIP)
To a circuit with the operating values:
U
= 10.5 ... 45 V DC; P
m
max
II 3 G Ex nA II T4/T6;
II 2/3 G Ex ib/nL IIC T4/T6;
II 2/3 G Ex ib/ic IIC T4/T6
U
= 45 V DC
m
U
= 45 V
n
L
= 400 μH
i
C
= 6 nF
i
XP CL I, DIV 1, GP ABCD T4/T6;
DIP CL II, III, DIV 1, GP EFG T4/T6;
IS CL I, II, III, DIV 1, GP ABCDEFG T4
CL I, Zone 0, AEx ia IIC T4;
CL I, Zone 1, AEX ib IIC T4
= T4: -40 ... +85 °C (-40 ... +185 °F)
a
T
= T6: -40 ... +60 °C (-40 ... +140 °F)
a
To "control drawing":
U
= 30 V, I
= 100 mA,
m
m
P
= 750 mW, L
= 400 μH, C
i
i
NI CL I, DIV 2, GP ABCD T4/T6;
NI CL I, Zone 2, GP IIC T4/T6;
S CL II, III, GP FG T4/T6;
NI CL I, DIV 2, GP ABCD T4/T6, NIFW;
NI CL I, Zone 2, GP IIC T4/T6, NIFW
NI CL II, III, DIV 2, GP FG T4/T6, NIFW
= T4: -40 ... +85 °C (-40 ... +185 °F)
a
T
= T6: -40 ... +60 °C (-40 ... +140 °F)
a
To "control drawing":
Um = 45 V, Li = 400 μH, C
CL I, DIV 1, GP ABCD T4 /T6;
CL II, DIV 1, GP EFG T4/T6
= T4: -40 ... +85 °C (-40 ... +185 °F)
a
T
= T6: -40 ... +60 °C (-40 ... +140 °F)
a
To "control drawing":
U
= 45 V
max
Technical data
11.8 Certificates and approvals
= 1.2 W
= 6 nF
i
= 6 nF
i
211