Page 1
Introduction Safety instructions Description SITRANS P Installing/mounting Pressure transmitter SITRANS P210 Connecting Commissioning Compact Operating Instructions Service and maintenance Technical specifications Dimension drawings Product documentation and support 7MF1566 09/2020 A5E03304388-AA...
Page 2
Note the following: WARNING Siemens products may only be used for the applications described in the catalog and in the relevant technical documentation. If products and components from other manufacturers are used, these must be recommended or approved by Siemens. Proper transport, storage, installation, assembly, commissioning, operation and maintenance are required to ensure that the products operate safely and without any problems.
• Devices/replacement parts should be returned in their original packaging. • If the original packaging is no longer available, ensure that all shipments are properly packaged to provide sufficient protection during transport. Siemens cannot assume liability for any costs associated with transportation damages.
The content reflects the technical status at the time of publishing. Siemens reserves the right to make technical changes in the course of further development.
Directive of the European Parliament and of the Council on the restriction of the use of certain hazardous substances in electrical and electronic equipment The standards applied can be found in the EC declaration of conformity for the device. SITRANS P210 Compact Operating Instructions, 09/2020, A5E03304388-AA...
Note Operation under special ambient conditions We highly recommend that you contact your Siemens representative or our application department before you operate the device under special ambient conditions as can be encountered in nuclear power plants or when the device is used for research and development purposes.
Page 9
• Connect the device with type of protection "Intrinsic safety" solely to an intrinsically safe circuit. • Observe the specifications for the electrical data on the certificate and/or in Technical specifications (Page 25). SITRANS P210 Compact Operating Instructions, 09/2020, A5E03304388-AA...
Page 10
Safety instructions 2.4 Use in hazardous areas SITRANS P210 Compact Operating Instructions, 09/2020, A5E03304388-AA...
It can be connected electrically with a connector fulfilling EN 175301-803-A (IP65) or a round plug M12 (IP67). The output signal amounts to between 4 and 20 mA. SITRANS P210 Compact Operating Instructions, 09/2020, A5E03304388-AA...
3 digits of the order number Order item Single part number Separator Decoding for year, month and day information Code Year 2010 2011 2012 2013 2014 2015 H (G) 2016 2017 2018 2019 2020 SITRANS P210 Compact Operating Instructions, 09/2020, A5E03304388-AA...
Page 13
Day of month 1st to 31st day Conforming to DIN EN 60062 The letter G is not permitted for new applications since it deviates from DIN EN 60062. It serves only for coding back. SITRANS P210 Compact Operating Instructions, 09/2020, A5E03304388-AA...
Page 14
Description 3.3 Layout of the nameplate SITRANS P210 Compact Operating Instructions, 09/2020, A5E03304388-AA...
Refer to the information in Technical specifications (Page 25). Note Material compatibility Siemens can provide you with support concerning selection of sensor components wetted by process media. However, you are responsible for the selection of components. Siemens accepts no liability for faults or failures resulting from incompatible materials.
• The overload limit must not be exceeded. • Connect the devices with fixed cable installation. • High-frequency electromagnetic interferences may result in measurement deviations of up to 1%. SITRANS P210 Compact Operating Instructions, 09/2020, A5E03304388-AA...
(process connection) or the ground conductor of the plug. Exception: It may be permissible to omit connection of the equipotential bonding for devices with type of protection "Intrinsic safety Ex i". SITRANS P210 Compact Operating Instructions, 09/2020, A5E03304388-AA...
Page 18
The device should only be supplied with limited energy according to UL61010-1 Second Edition, Section 9.3 or LPS (Low Power Supply) in accordance with UL60950-1 or Class 2 in accordance with UL1310 or UL1585. SITRANS P210 Compact Operating Instructions, 09/2020, A5E03304388-AA...
Connecting with current output and fast-fit cable gland Connecting with voltage output and fast-fit cable gland Version with explosion protection: 4 to 20 mA The grounding connection is conductively bonded to the transmitter enclosure. SITRANS P210 Compact Operating Instructions, 09/2020, A5E03304388-AA...
Page 20
Connecting 5.2 Electrical connections Connecting with current output and connector in accordance with EN 175301 (Ex) Connecting with current output and connector M12x1 (Ex) SITRANS P210 Compact Operating Instructions, 09/2020, A5E03304388-AA...
• Before commissioning take the effect on other devices in the system into account. Correction of zero point and span The transmitter is preset to the specific measuring range by the manufacturer. An additional setting is not possible. SITRANS P210 Compact Operating Instructions, 09/2020, A5E03304388-AA...
Page 22
Commissioning 6.1 Correction of zero point and span SITRANS P210 Compact Operating Instructions, 09/2020, A5E03304388-AA...
WARNING Impermissible repair of explosion protected devices Risk of explosion in hazardous areas • Repair must be carried out by Siemens authorized personnel only. WARNING Use of a computer in a hazardous area If the interface to the computer is used in the hazardous area, there is a risk of explosion.
Service and maintenance 7.4 Disposal Required forms • Delivery note • Return document (http://www.siemens.com/processinstrumentation/returngoodsnote) with the following information: – Product (item description) – Number of returned devices/replacement parts – Reason for returning the item(s) • Decontamination declaration (http://www.siemens.com/sc/declarationofdecontamination) With this declaration you warrant "that the device/replacement part has been carefully cleaned and is free of residues.
Effect of ambient temperature • Start of scale value and measuring span • 0.25%/10 K of full-scale value • 0.5%/10 K of full-scale value for a measuring range of 100 ... 400 mbar SITRANS P210 Compact Operating Instructions, 09/2020, A5E03304388-AA...
Page 26
-40 ... +125 °C < 100 bar (‑40 ... +257 °F; < 1 450 psi), can be used for drinking water Non-wetted parts material Stainless steel, Mat. No. 1.4404 (SST 316 L) • Enclosure SITRANS P210 Compact Operating Instructions, 09/2020, A5E03304388-AA...
≤ 100 mA; P ≤ 0.75 W circuits with maximum values Effective internal inductance and capacitance for = 0 nH; C = 0 nF versions with plugs according to EN 175301-803-A and M12 plugs SITRANS P210 Compact Operating Instructions, 09/2020, A5E03304388-AA...
– "Download as html5, only PC": Open or save the manual in the HTML5 view on your PC You can also find manuals with the Mobile app at Industry Online Support (https:// support.industry.siemens.com/cs/ww/de/sc/2067). Download the app to your mobile device and scan the device QR code. Product documentation by serial number Using the PIA Life Cycle Portal, you can access the serial number-specific product information including technical specifications, spare parts, calibration data, or factory certificates.
Additional information on our technical support can be found at Technical Support (http:// www.siemens.com/automation/csi/service). Service & support on the Internet In addition to our technical support, Siemens offers comprehensive online services at Service & Support (http://www.siemens.com/automation/serviceandsupport). Contact If you have further questions about the device, contact your local Siemens representative at Personal Contact (http://www.automation.siemens.com/partner).