Measurement In A Closed Container - Siemens Sitrans P500 Operating Instructions Manual

Pressure transmitter
Table of Contents

Advertisement

4.6.3.2

Measurement in a closed container

When taking measurements in a closed container without or with little condensation
formation, the negative pressure line is not filled.
h
U
h
O
p
Measurement assembly on a closed container:
no or little condensation separation
When taking measurements in a closed container with strong condensation formation, you
must fill the negative pressure line (mostly with the condensation of the measured medium)
and install a condensation pot.
For measurement on open vessels there is no need to connect the minus pipework, since
the minus side is open to the atmosphere.
SITRANS P500 with HART
Operating Instructions, 12/2014, A5E02344528-09
Start of scale value
Full scale value
Pressure
Formula:
Start of scale value
Δp
Full scale value
Δp
Δp
Start of scale value to be adjusted
MA
Δp
Full scale value to be adjusted
ME
ρ
Density of the measured medium in
the container
g
Acceleration due to gravity
Install
4.6 "Level" installation
= ρ · g · h
MA
U
= ρ · g · h
ME
O
43