Before Returning an Instrument to a Customer

Before Returning an Instrument to a Customer
When finished servicing an FieldFox analyzer, perform the following before returning the unit to the
customer:
• Set the Display Colors to Indoor and the display Brightness to 80% via
System (7) > Preferences > Preferences, etc.
• Perform an Operational Check as described in the following section.
• Connect and disconnect the external battery charger while the FieldFox is operating and confirm that
the power cord icon is displayed appropriately at the top of the display.
• Charge the battery to 80% or higher.
• Press
System > System Configuration > More > Date and Time
See http://na.tm.agilent.com/fieldfox/service/rtc_info.htm
for restrictions and details for setting time and date.
3-6
FieldFox Analyzer
.
N9927-90003