Technical data
11.4 Cable data
11.4
Cable data
Description
Cable for standard electrode or coil
Electrode cable, double shielded
Cable kit with standard coil cable and electrode cable double
shielded (also available as low noise cable for MAG 1100
sensor )
Technical data
Basic data
Ambient temperature
Cable parameter
11.5
Cable requirements
Basic data
Max. cable loop resistance
Cable glands on sensor and
transmitter
68
No. of conductors
Sqr. area
Screen
Color code
Outside color
Ext. diameter
Conductor
Isolation material
Flexible installation
Non-flexible installtaion
Capacity
Inductance
L/R
No. of conductors
Min. sqr. area
Screen
Max. capacitance
Media temperature:
< 100 °C
> 200 °C
M20x1.5 gland - Cable ø 5 ... 13 mm (0.20 ... 0.51 inches)
½ NPT gland - cable ø 5 ... 9 mm (0.20 ... 0.35 inches)
Standard electrode
cable
(electrode/coil)
3
1.5 mm
2
Yes
Brown, blue, black
Grey
7.8 mm
Flexible CU
PVC
-5 ... +70°C
(23 ... 158°F)
-30 ... +70°C
(-22 ... 158°F)
161.50 pF/m
0.583 μH/m
43.83 þH/Ω
Coil
cable
2
0.5 mm
2
Yes
N/A
40 Ω
6 Ω
SITRANS F M MAG 5000/6000
Operating Instructions, 01/2010, SFIDK.PS.026.G1.02
Special cable
(electrode)
3
0.25 mm
2
Double
Brown, blue, black
Grey
8.1 mm
Flexible CU
PVC
-5 ... +70°C
(23 ... 158°F)
-30 ... +70°C
(-22 ... 158°F)
N/A
N/A
N/A
Electrode
cable
3
0.2 mm
2
Yes
350 pF/m
N/A
N/A