Siemens SITRANS F M MAG 5000 Operating Instructions Manual page 62

Electromagnetic flowmeters
Hide thumbs Also See for SITRANS F M MAG 5000:
Table of Contents

Advertisement

Troubleshooting/FAQs
Symptom
Output
signals
Maximum
Measuring approx.
50%
Loss of totalizer data
OK
##### Signs in
OK
display
60
Error
Cause
code
F71
Loss of internal data
W30
Flow exceeds 100% of Qmax.
W21
Pulse overflow
Volume/pulse too small
Pulse width too large
Missing one electrode
connection
W20
Initializing error
Totalizer roll over
Remedy
Replace MAG 5000/6000
Check Q
(Basic Settings)
max
Change volume/pulse
Change pulse width
Check cables
Reset totalizer manually
Reset totalizer or increase totalizer
unit
SITRANS F M MAG 5000/6000
Operating Instructions, 01/2010, SFIDK.PS.026.G1.02