Accuracy - Siemens SITRANS F M MAG 5000 Operating Instructions Manual

Electromagnetic flowmeters
Hide thumbs Also See for SITRANS F M MAG 5000:
Table of Contents

Advertisement

Technical data

11.2 Accuracy

11.2
Accuracy
For accuracy reference conditions, please see below.
[±% E]
3,0
2,8
2,6
2,4
2,2
2,0
1,8
1,6
1,4
1,2
1,0
0,8
0,6
0,4
0,2
0,0
0
1
0
3.3
Figure 11-1
MAG 5000 with MAG 1100, MAG 1100 F, MAG 5100 W, MAG 3100 and MAG 3100 P
[±% E]
1,4
1,2
1,0
0,8
0,6
0,4
0,2
0,0
0
1
0
3.3
Figure 11-2
MAG 6000 with MAG 1100 (not PFA), MAG 1100 F (not PFA), MAG 5100 W, MAG 3100 and MAG 3100 P
64
V: Flow velocity
E: Meter uncertainty as a percentage of measured value
v ≥ 0.1 m/s (0.3 ft/s) --> E: ±0.4 ± 1 mm/s *
v < 0.1 m/s (0.3 ft/s) --> E: ±(0.25/v) % of measured value
* MAG 5100 W with DN 350 ... 2000
v ≥ 0.1 m/s (0.3 ft/s) --> E: ±0.4 ± 2 mm/s
v < 0.1 m/s (0.3 ft/s) --> E: ±(0.25/v) % of measured value
2
3
4
6.6
9.9
13
V: Flow velocity
E: Meter uncertainty as a percentage of measured value
v ≥ 0.1 m/s (0.3 ft/s) --> E: ±0.2 ± 1 mm/s *
v < 0.1 m/s (0.3 ft/s) --> E: ±(0.125/v) % of measured value *
* MAG 5100 W (Order No. 7ME652...) with DN 350 ... 1200
v ≥ 0.1 m/s (0.3 ft/s) --> E: ±0.2 ± 2.5 mm/s
v < 0.1 m/s (0.3 ft/s) --> E: ±(0.25/v) % of measured value
2
3
4
5
6.6
9.9
13
16
5
6
7
8
16
20
23
26
6
7
8
20
23
26
9
10
11
30
33
36
9
10
11
12 [m/s]
30
33
36
39
SITRANS F M MAG 5000/6000
Operating Instructions, 01/2010, SFIDK.PS.026.G1.02
12
[m/s]
39
[ft/s]
[ft/s]

Hide quick links: