Commissioning

In this chapter it is described how to commission the device via the local user interface (LUI).
The display is described in details in section Local user interface (Page 38).
Furthermore, the following functions are described in details:
● Changing password (Page 40)
● Changing basic settings (Page 40)
● Changing operator menu setup (Page 42)
● Changing language (Page 43)
Detailed diagrams concerning the specific menu are shown in appendix menu diagrams
(Page 73).
For factory settings, see Factory settings (Page 91).
6.1

MAG 5000/6000 Blind

Note
Does not have a display. All factory settings will be uploaded from the SENSORPROM
after power-up.
For sensor dependent factory settings, see Appendix B (Page 93).
Changing settings
If other settings are required, a standard transmitter with display and similar power supply
can be used.
1. Unscrew and remove MAG 5000/6000 Blind.
2. Mount standard MAG 5000/6000 transmitter.
3. Change required settings via display and keypad.
4. Remove standard transmitter and remount Blind transmitter.
5. Fasten screws holding transmitter.
New settings stored in SENSORPROM
SITRANS F M MAG 5000/6000
Operating Instructions, 01/2010, SFIDK.PS.026.G1.02
All changed data will be stored in SENSORPROM
memory unit.
®
memory unit will be uploaded in blind transmitter.
®