Physical Data; Specifications And Recommended Operational Parameters - Thermo Scientific Acclaim HILIC-10 Product Manual

Analytical column
Table of Contents

Advertisement

1
– Introduction

1.2 Physical data

Bonding Chemistry:
Silica Substrate:
Analytical
Guard
(requires holder
P/N 069580)

1.3 Specifications and Recommended Operational Parameters

Shipping solution:
Storage solution:
pH Range:
Temperature Range:
Particle
Size
3 µm
5 µm
1.9 µm
Thermo Scientific
065339-02
Proprietary covalently bonded hydrophilic layer
Spherical, high-purity
Particle Type
3 µm particle size
120 Å pore size
2
300 m
/g surface area
5 µm particle size
120 Å pore size
2
300 m
/g surface area
1.9 µm particle size
175 Å pore size
2
200 m
/g surface area
3 µm particle size
120 Å pore size
2
300 m
/g surface area
90/10 acetonitrile/100 mM ammonium acetate
90/10 v/v acetonitrile/10-50 mM ammonium acetate or 100%
acetonitrile
pH 2 – 8
< 70° C
Maximum
Column Dimension
Recommended Pressure
4.6 x 150 mm
8500 psi
3.0 x 150 mm
8500 psi
2.1 x 150 mm
8500 psi
4.6 x 250 mm
6000 psi
4.6 x 150 mm
6000 psi
2.1 x 150 mm
6000 psi
2.1 x 150 mm
10,000 psi
2.1 x 250 mm
15,000 psi
Product Manual for Acclaim HILIC-10 Columns
Column Dimensions
4.6 x 150 mm
3.0 x 150 mm
2.1 x 150 mm
4.6 x 250 mm
4.6 x 150 mm
2.1 x 150 mm
2.1 x 150 mm
2.1 x 250 mm
3.0 x 10 mm
4.6 x 10 mm
2.1 x 10 mm
Typical Flow Rate
0.8 – 1.5 mL/min
0.4 – 0.8 mL/min
0.2 - 0.4 mL/min
0.8 – 1.5 mL/min
0.8 – 1.5 mL/min
0.2 – 0.4 mL/min
0.2 – 0.4 mL/min
0.2 – 0.4 mL/min
P/N
074257
074258
074259
079693
079694
079695
074260
075709
074261
074262
074263
Page 8 of 23