b) Entity parameters and connections of input or measurement
b.1) Voltage (V) Input:
b.2) Current (mA) Input:
U
= 30 V
i
I
= 100 mA
i
P
= 750 mW
i
C
= 0 nF
C
i
L
= 400 µH
L
i
Fig. 03 – V Input
U
= 30 V
i
I
= 100 mA
I
i
P
= 750 mW
P
i
C
= 0 nF
C
i
L
= 400 µH
L
i
Fig. 04 – mA Input
U
= 5.4 V
o
I
= 6 µA
o
P
= 9 µW
o
= 21.9 nF
o
= 1.75 mH
o
U
= 3.5 V
o
= 5.9 mA
o
= 8.7 mW
o
= 24.6 nF
o
= 1.35 mH