Scale Function (Out) - Presys PC-507-IS Technical Manual

Pressure calibrator
Table of Contents

Advertisement

To disable SCALE or CAL functions, just select NO option from the menu
below and press ENTER.
OUTPUT has the SCALE, CONV and NO options, described as follows.
The output type must be previously configured; otherwise the SELECT
OUTPUT FIRST message will be displayed. In this case, press C/CE to go back to
the main menu and select the output type.

3.9.3. SCALE Function (OUT)

Scale function of the calibrator output allows it to simulate the operation of a
transmitter. Transmitter input is made directly by keyboard, and it can generate
voltage or current as output signal.
SCALE output function relates to the output signal generated by the calibrator
to the value shown on display, according to the example shown below.
The scaled display indication (#) may represent any engineering unit, such as
3
m/s, m
/s, % etc.
SCALE
CAL
 NO
Page 43