4.2. Output Adjustment

Select the corresponding mnemonic, choose the setpoint explained as
follows, measure the signal generated by the calibrator and store this value as
detailed in the following tables.
For adjustment of outputs, the display shows three values:
First Field (1) is the value of the setpoint as a percentage (%) of the
output range required by the user and it is selected by pressing key "0" (zero).
Second Field (2) is the value measured by the calibrator expressed in
percentage (%) of the output range. Before providing the 1
points, one must wait until this value stabilizes.
Third Field (3) is the value entered by the user after the output has been
measured and the values corresponding to the two setpoints have been stored: 1
nd
point (key 1), 2
point (key 2) and 3
V output
Single range
mA output
Single range
SP = 50,000%
(2)
49,999
5,00000
rd
point (key 3).
st
1
point
SP = 10.000%
st
1
point
SP = 10.000%
(1)
(3)
st
, 2
nd
2
point
SP = 60.000%
nd
2
point
SP = 60.000%
Page 54
nd
rd
and 3
adjustment
rd
3
point
SP = 80.000%
rd
3
point
SP = 80.000%