Available Power Supply (Tps) - Presys PC-507-IS Technical Manual

Pressure calibrator
Table of Contents

Advertisement

b.3)
Current (mA) output passive mode:
CAUTION: Always connect the terminal mA OUT(-) to the
positive and the terminal mA IN (-) to the negative
in order to not invalidate the protection mode.

3.6. Available Power Supply (TPS)

The Calibrator has a 12V@20mA regulated power supply (TPS), with
short-circuit protection (current limited to 30 mA).
U
= 30 V
U
i
I
= 100 mA
i
P
= 750 mW
P
i
C
= 0 nF
i
L
= 400 µH
i
Fig. 11 – mA Output passive mode
U
= 17.1 V
o
I
= 73 mA
o
P
= 309 mW
o
C
= 120 nF
o
L
= 2.93 mH
o
Fig. 12 – Power Supply (TPS)
= 17.1 V
o
I
= 31 mA
o
= 27 mW
o
C
= 33 nF
o
L
= 1 mH
o
Page 28