3.8. Special Programming

When PRG is selected, the display shows:
This option allows the selection of special programming features for INPUT or
OUTPUT. INPUT is provided with FILTER, DECIMAL and SPEED options. OUTPUT
is provided with STEP and RAMP options.

3.8.1. FILTER Programming

The value of this parameter (in seconds) configures the time constant of a first
order digital filter coupled with the selected input. When the filtering of the measured
signal is not required, just set this parameter as 0 (zero).

3.8.2. DECIMAL Programming

The value of this parameter (0, 1, 2, 3 or DEFAULT) indicates the number of
decimals that the value measured at the input will be displayed.
Note: DEFAULT corresponds to the maximum decimals the calibrator may
display in an input measurement, in accordance with its resolution.