HP 8753E Programming And Command Reference Manual page 112

Hp-ib network analyzer, including option 011
Hide thumbs Also See for HP 8753E:
Table of Contents

Advertisement

These 3 commands control the warning beeper, causing it to sound if the
indicated condition occurs:
These commands set the open capacitance values of an open circuit
while it is being defined as a calibration standard:
These commands set the power meter calibration factor corrections for
the particular sensor used. Sensor B is only valid for the HP 438A which
has two input channels:
HP-IB Frogr amming and Command Reference Guide
The completion of functions such as save,
done with calibration standard, and data
trace saved.
A limit test failure.
The generation of a warning message.
Blanks the display.
Measures and displays B/R on the active
channel.
Range: f10k(10-15 F)
Range: 6 1 Ok(1 Omz7 F)
Range: ~10k(10-s6 F)
Range: 3dOk(10-45 F)
Accepted for compatibility with the HP
8510, where its function is to begin a
calibration sequence.
Sets the calibration factor. Range: 0 to
200%.
Selects the frequency for the calibration
factor correction. Range: stimulus
Edits the sensor A calibration factor table.
Edits the sensor B (HP 438A only)
calibration factor table.
range.