Specifications
7.2

7.2.1 Physical

Dimensions
(WxDxH)
Weight

7.2.2 Electrical

Power
Requirements
Fuse
INPUTS
(VALVE A,
VALVE B,
SSV)

7.2.3 Communication

Switching Valve
A and B
Solvent
Selection valve
Loading Pump
7.3
Inert Version
(factory
installed)
Flow Passage
Diameter
7-2

General

185 mm (7.3 in) x 350 mm (13.8 in) x 430 mm (16.9 in).
16 kg (35 lbs.).
90 - 260 V, 47 - 62 Hz, 100 VA
2x Slow-Blow, 2A
Input resistance:
max. Voltage 'Input inactive'
min. Voltage 'Input' active
absolute maximum ratings:
min. voltage:
max. voltage:
Note: Maximum ratings are the extreme limits the input pin can be exposed
to without causing permanent damage. Do NOT connect voltages
lower/higher than those limits.
by RS-232 serial communication
or
by relay, open collector or TTL outputs
(e.g., from FAMOS or UltiMate Micropump)
by RS-232 serial communication
or
by relay, open collector or TTL outputs
(e.g., from FAMOS or UltiMate Micropump)
by RS-232 serial communication network

Options

all fluidic pathways made out of inert materials,
i.e. PAEK, PEEK, PTFE, ceramic, titanium
0.25 mm
D850R3
10 kΩ
0.5 V
2.0 V
-0.3 V
5,3 V