Using The Event Outputs Or The Famos Auxiliaries - LC Packings SWITCHOS II User Manual

Advanced microcolumn switching unit
Table of Contents

Advertisement

Installation and Getting Started
2.4
If no additional COM port is available or if the Switchos Advanced Microcolumn
Switching Unit is to be installed with a different or third vendor software
package, the switching valves on the Switchos can be controlled by:
a)
the event outputs of the Switchos Loading Pump.
b)
the Auxiliaries (P5) outputs of the FAMOS Microautosampler.
To connect the INPUTS connector:
a) Valves are controlled by the Loading Pump
Connect the INPUTS connector to the EVENTS outputs of the loading pump
using the connection cable P/N 160172 presented in FIGURE 2-4.
EVENT3
GROUND
EVENT6
GROUND
{
EVENT7
{
EVENT8
FIGURE 2-4 Cable to control Switchos by the UltiMate System
b) Valves are controlled by the FAMOS Microautosampler
Connect the INPUTS connector of the Switchos, the P5 (AUXILIARIES)
connector of the FAMOS Microautosampler and the START IN input of the
UltiMate UV Detector using the connection cable P/N 160171 presented in
FIGURE 2-5.
FAMOS
(P5 Auxiliaries)
FIGURE 2-5 Cable to control Switchos by the FAMOS P5 Connector
2-6

Using the Event Outputs or the FAMOS Auxiliaries

Switchos
Events
to MS or another device
D850R3
Micropump Switchos II
EVENT3
EVENT6
EVENT7
EVENT8
Switchos
INPUTS
FAMOS
Switchos II
AUX1
Valve A
AUX2
Valve B
AUX3
Switchos
AUX4
START IN
(INPUTS)
UltiMate
UV Detector
(START IN)
User's Manual Switchos
Valve A
Valve B
SSV
Trigger
SSV