LC Packings SWITCHOS II User Manual page 70

Advanced microcolumn switching unit
Table of Contents

Advertisement

Maintenance and Troubleshooting
TABLE 6-2 Troubleshooting the Switching Valves
The Switchos does not
respond to any signal
applied to the INPUTS
connector
The switching valves do not
respond when pressing the
controls (push buttons on
front/rear panel)
The Solvent Selection valve
moves to the next solvent
channel, and after a short
moment it moves back to
channel A (REMOTE control
mode)
The valve A (B) switches to
position 1-2 as soon as the
instrument is set into
REMOTE control mode
LEDs 1-2 and 10-1 of Valve
A (B) are switched off or
switched on at the same
time
LED A-D illuminated at the
same time
Note: It is important to note that many analytical problems are be caused by
external influences such as the temperature and or the analysis of light sensitive
samples and are not instrument related. When troubleshooting, make certain that
the application was running in an acceptable fashion before the problems were
observed and that nothing has been changed in the application. If you are making
a change in the protocol, change a single parameter at a time and ensure that the
effect of each change is well understood.
6-12
The Switchos is in LOCAL
mode, e.g. the LEDs in the
push buttons are illuminated
The Switchos is in REMOTE
control mode, e.g. the LEDs
in the push buttons are not
illuminated
The signal applied to the
SSV input has been active
for longer 1.5 sec
At the present time there is
no active input signal and
the valve is forced into
position 1-2
The Switching Valve A (B)
does not switch...
due to valve sticking, ...
clamp ring slippage,..
other positioning problems
The Solvent Selection Valve
does not switch...
due to valve sticking, ...
clamp ring slippage,..
other positioning problem
D850R3
Change control mode to
REMOTE (REMOTE/LOCAL
switch on the rear panel)
Change control mode to
LOCAL (REMOTE/LOCAL
switch on the rear panel)
Apply a shorter signal
(< 1.5 seconds)
Change the setting of the
instrument controlling
Switchos according to the
needs of the application
Switch OFF the instrument
and ON again to reset the
error
Clean Valve
Tighten clamp ring
Contact LC Packings
Switch OFF the instrument
and ON again to reset the
error
Contact LC Packings
Contact LC Packings
Contact LC Packings
User's Manual Switchos II