2.5
The electrical connections that are required depend on the nature of the
instrumentation and the desired application. In this section, we provide general
information about how the instrument can be interfaced to equipment from other
manufacturers. The user should also refer to the documentation provided with
this equipment. The Switchos is controlled via the INPUTS connector on the rear
panel and the keypad of the loading pump.
Caution: If you are interfacing the Switchos to instrumentation from other
CAUTION
manufacturers, ensure that input voltages and output voltages are within the
ranges indicated in the specifications (Chapter 7). If you have any questions,
please call LC Packings. The LC Packings warranty will not be valid if the
Switchos is damaged due to interfacing of the system to instrumentation from
third parties.

2.5.1 INPUTS Connector

The positions of the switching valves A and B and the Solvent Selection Valve
(SSV) are controlled by the INPUTS connector (FIGURE 2-6). Contact closure
(relay), TTL and open collector outputs can drive the Switchos inputs.
Caution: Check to make certain that the maximum ratings of the outputs
CAUTION
matches with the requirements of the Switchos Advanced Microcolumn
Switching Unit (Chapter 7).
8
15
FIGURE 2-6 Pin-Out INPUTS Connector
To control the switching valves A and B and the SSV:
Connect two outputs of the controlling instrument to the Switchos input pins
VALVE A and VALVE B of the INPUTS connector (FIGURE 2-7).
Connect the common terminals or the ground terminals of the controlling
instrument to the corresponding GROUND pins of the INPUTS connector
(FIGURE 2-7).
Connect the SSV pin and the corresponding GROUND in the same way to
another output of the controlling instrument (FIGURE 2-7).
Controlling
Instrument
Relay
A) Contact Closure (Relay) Output
FIGURE 2-7 Different Output Configurations of the Controlling Instrument
User's Manual Switchos

Installing the Switchos with Other HPLC Systems

INPUTS
1
9
VALVE A
VALVE B
SSV
GROUND
D850R3
Installation and Getting Started
PIN #
SIGNAL
1
VALVE A
2
GROUND
4
VALVE B
5
GROUND
7
SSV
8
GROUND
Controlling
Instrument
B) TTL / Open Collector Output
VALVE A
VALVE B
SSV
GROUND
2-7