LC Packings SWITCHOS II User Manual page 35

Advanced microcolumn switching unit
Table of Contents

Advertisement

Dimension
50 cm x 130 µm I.D.
100 cm x 130 µm I.D.
b) Connect port 6 of the injection valve of the FAMOS Microautosampler to the
port 1 of valve A of the Switchos II as described in item a).
c) Connect the outlet flow of the UltiMate System (e.g. on the left side panel or
from the upper T-Piece of the flow splitter) to port 4 of valve A of the
Switchos. Use the appropriate tubing for your application.
Application (I.D.)
Capillary LC (50 µ m)
Nano LC (20 µm)
d) Connect port 3 of valve A of the Switchos to the UltiMate column bulkhead or
directly to the micro column located in the UltiMate column compartment. Use
the appropriate tubing for your application.
Application (I.D.)
Capillary LC (50 µm)
Nano LC (20 µm)
e) Connect ports 6 and 10 of valve A of Switchos to waste (e.g. using 200 -
500 µm I.D. PTFE tubing).
f) Connect the trap column between ports 2 and 5 of valve A of the Switchos II,
using the appropriate tubing supplied with the trap column.
A typically loading solvent is 100% water with 0.05% TFA. Fill solvent bottle A
with this solvent. Fill bottle B with either the same solvent or a mixture of
methanol/water (1:1), bottles C and D may be left empty).
After pre-concentrating the sample on the trap column, the valve A of Switchos
will be switched into position 10-1 to elute the sample (FIGURE 2-13).
FAMOS
port 6
FAMOS
port 1
FIGURE 2-13 Eluting Sample adsorbed on the Trap Column (in Backflush Mode)
User's Manual Switchos
Standard Version
P/N 160180
P/N 160181
Standard Version
P/N 161479
P/N 160178
Standard Version
P/N 161480
P/N 160179
Switchos
Valve A
Trap column
D850R3
Installation and Getting Started
Inert Version
P/N 160180
P/N 160181
Inert Version
P/N 161261
P/N 161259
(1) use PEEK fingertight fittings only
Inert Version
P/N 161262
P/N 161260
(1) use PEEK fingertight fittings only
Analytical microcolumn
Micro flow
to UltiMate
Waste
1
10
2
9
3
Valve A
presented
in position
8
4
10-1
5
7
6
Micro flow
from UltiMate
Waste
(1)
(1)
(1)
(1)
2-13