Software Control
4.2.3
If serial control is not possible (e.g. no free COM port available), the Switchos
loading pump can be used to control the valve positions. The switching valves
are then controlled by the event outputs of the Switchos II loading pump
(TABLE 4-1) and EVENT8 can be used to trigger another device (e.g. a mass
spectrometer).
TABLE 4-1 Control Outputs
Switchos II Input
Valve A
Valve B
SSV
Trigger
Note: The EVENT number as well as the type are indicated on the rear panel of
the Micropump (e.g. EVENT 1 'TTL', EVENT 3 'OC', etc.). Make certain that the
proper EVENT output is linked to the corresponding signal name of
CHROMELEON (e.g. 'Valve_A' should control EVENT3 of the Micropump, 'SSV'
should control EVENT7, etc.).
To control the unit by the event outputs of the Loading Pump:
a) Select the Relays tab of the LC Packings UltiMate/Switchos box (FIGURE 4-4).
FIGURE 4-4
b) Check the outputs 'SSV', 'Valve_A', and 'Valve_B', check 'Switchos_Relay_2'
if you want to trigger any instrument by the EVENT 8 output.
Note: The default signal names corresponds to the names used in the
(predefined) panels and should not be changed. If you modify the signal names,
they may not be recognized and some functions may not work properly.
4-4

Switchos II controlled by Event Outputs

Loading Pump
EVENT3
EVENT6
EVENT7
EVENT8
The UltiMate Configuration Box – Relays Tab
D850R3
Default CHROMELEON Signal Name
Valve_A
Valve_B
SSV
Switchos_Relay2
User's Manual Switchos II