Maintenance and Troubleshooting

6.3.2 Replacing the High Pressure In-line Filter

The high pressure in-line filter serves to remove particulate matter from the
mobile phase and is considered as a consumable item. If excessive pressure is
required to deliver the mobile phase (indicated on the pump display), it is
probable that the filter should be replaced.
To replace the high pressure in-line filter:
e) Remove the capillary (item 1, FIGURE 6-2) between the purge valve and the
filter unit.
1
FIGURE 6-2. High Pressure Inline Filter
f)
Use the socket driver (P/N 160073) or a wrench (11 mm) to remove the filter
holder (item 2, FIGURE 6-2).
g) Remove the filter and place a new filter into the holder (FIGURE 6-3).
FIGURE 6-3. In-Line Filter in Filter Holder
h) Re-mount the filter holder.
Caution: Make certain that the filter remains in the filter holder when screwing
CAUTION
the holder in and do not overtighten the filter holder. Overtightening could result
in distortion of the filter which may lead to leakage and/or premature failure of
the component.
6-4
2
Filter
Description
Filter
Standard (color: black)
Inert
(color: blue)
Filter Frit Holder
Standard (stainless steel)
Inert
(titanium)
D850R3
P/N
160072
161107
163021
162108