Attention
:
Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches)
to any part of the plusoptiX 12, including cables specified by the manufacturer.
Otherwise, degradation of the performance of this equipment could result. The
plusoptiX 12 is not compatible to HF SURGICAL EQUIPMENT.
Recommended safety distances between portable and mobile HF telecommunication devices and the "plusoptix 12"
The "plusoptiX 12" is intended for use in an electromagnetic environment with controlled HF disturbance variables. The customer
or user of the "plusoptiX 12" can contribute to avoiding electromagnetic disturbances, by complying with the minimum distance
between portable and mobile HF telecommunication devices (transmitters) and the "plusoptiX 12" – depending on the output
power of the communication device, as stated below.
Maximum output power of the
transmitter in W
0.01
0.1
1
10
100
For transmitters, whose maximum nominal output is not stated in the above table, the recommended safety distance in meters
(m) can be calculated using the equation from the relevant column, wherein P equals the maximum nominal output of the
transmitter in watts (W) as per information by the transmitter's manufacturer.
COMMENT 1 At 80 MHz and 800 MHz, the higher frequency range applies.
COMMENT 2 These guidelines may not be applicable in all cases. The distribution of electromagnetic quantities is influenced by
absorption and reflections from buildings, objects and people.
Safety distance depending on the transmission frequency (in meters)
150 kHz - 80MHz
d = 0.35 √ P
0.035
0.11
0.35
1.11
3.5
Seite 25/25
80 MHz - 800MHz
d = 0.35 √ P
0.035
0.11
0.35
1.11
3.5
800 MHz - 2,7GHz
d = 0.7 √ P
0.07
0.22
0.7
2.21
7