NOTE:
During programming you may backspace by using
Pressing
Pressing
Pressing

TPN Infusion

Total Parenteral Nutrition
The pattern of the program is a trapezoid.
TPN Trapezoid Program
Rate
Max Rate
On a TPN Program the operator may set the total volume which corresponds to the size of the bag to be
infused, the total time, and the up and down time. Program operation will start from 2 ml/h and will
increase to the maximal rate. During the flat session (T2) the rate will not change. During the last part of
the program (T3), the program will start to decrease the infusion rate according to the set taper down
time.
Up and down time can be set to zero. A TPN program with up and down time set to zero, is a continuous
program.
NOTE:
Before operating the pump in TPN program, make sure that the IV bag contains an
additional 5 ml of total volume to be infused. This ensures adequate volume for KVO during End
Program. If Delay Before Start is set, you must include enough volume for the delay period in
addition to the 5 ml KVO volume.
1.
Turn the pump on by pressing and holding the
Warning:
if disabled. Using the pump with air in line detector off may cause an embolism resulting death or
paralysis. Switching the air detector off is not recommended and should be used only when
absolutely necessary.
VET-PRO VIP 2000™ Color Vision Operator Manual
once clears the last entered digit.
again clears the next number.
when the value is empty displays the original value or the previous screen.
- This mode delivers an infusion while automatically tapering up and down.
T1
T2
If air sensor is off, please contact your technician in order to enable the air sensor
T1 = Time UP
T2= Time FLAT
T3
T3= Time DOWN
Time
until the Self-Test screen appears.
4 – Operation
.
62