4 – Operation
Before Operating the Pump
Before attaching the system to a patient, run the following tests to verify that all indicators and alarms
work properly.
When an alarm is activated the following occurs:
Infusion ceases
An audible alarm sounds
An alarm message appears on the display
The LED changes from green to red
NOTE:
All tests should be performed in the Continuous program.
Pump Operation Test
1.
Click the pump into a Charger connected to AC power. Verify that the CHARGE indicator is on,
Red or Green light (on charger).
2.
Insert an administration set and close the door. Press the door until a click is heard.
NOTE:
For instructions on how to load and prime an administration set refer to sections
"Loading the Administration Set" & "Priming the Line".
3.
Press
until the Self-Test screen appears. If air sensor is off, press
4.
The program selected and its status are displayed. During the Self-Test a long beep will be
heard. This procedure verifies that the pump's acoustic and visual features are working properly.
After self test is completed, the display will show the last rate setting.
Press
to prime the IV line.
5.
Warning:
patient!
VET-PRO VIP 2000™ Color Vision Operator Manual
Ensure the set is not connected to a
to confirm.
Continuous
10:00
Prime
Warning !
Disconnect Patient
Press
key to Start
4 – Operation
80%
███■
-
21