Manual Prime Operation; Operation Steps - CME VET-PRO VIP 2000 Operator's Manual

Veterinary infusion pump system
Hide thumbs Also See for VET-PRO VIP 2000:
Table of Contents

Advertisement

Press
once.
3.
Warning:
4.
Enter the volume to prime the set and press
begin priming procedure. You may stop priming at any
time by pressing
5.
The display screen will show a graph indicating the prime
operation.
6.
When priming is completed make certain that all air is removed from the IV line.
7.
Connect IV line to the patient and press

Manual Prime Operation

1.
In any case (before operation or after air in line alarm) a prime operation is needed, the user can
prime the set manually by using the manual priming valve, if assembled in the administration set
in use at that moment.
2.
Disconnect IV tubing from patient.
3.
Hold the tubing in a vertical position to maintain gravitation pressure. Press the priming valve
button continuously until all trapped air is primed. Release the priming valve button to continue
the anti free flow protection while infusing.

Operation Steps

The following are the steps for starting the infusion procedure. For detailed instructions regarding each step,
refer to the relevant sections in the Operation chapter.
VET-PRO VIP 2000™ Color Vision Operator Manual
Ensure the set is not connected to the patient!
.
Continuous
Prime
Warning !
Disconnect Patient
Press
Continuous
to
Prime Volume
5
mL
Press
Continuous
Priming
██████████████
1
to resume program.
4 – Operation
80%
███■
10:00
-
key to Start
80%
███■
10:00
-
key to Start
80%
███■
10:00
-
5mL
30