12.
Interval Time Screen
Do one of the following:
 Press
 Enter the time between the start of each dose and
press
.
13.
KVO Rate Screen
Do one of the following:
 Press
 Enter the KVO Rate (from 0.1 to 5 ml/h) and press
14.
Accumulation Screen
Do one of the following:
 Press
infusion.
 Press
to go back and adjust program parameters.
NOTE:
If pump is in Delay Mode, enter Start Time (24 Hour Clock) and Start Date. For further
information regarding the Delay Option, refer to
Delay Program section.
15.
Press
to Start Infusion.
16.
During program operation, the pump displays the rate in which the program is operating.
Resume Infusion, Start New Bag, or Restart Infusion
When an Intermittent infusion is unlocked , stopped and the pump is re-started, the user has the following
options:
(1) Resume: Resume the original infusion exactly where it left off
(2) New Bag: Resume the original infusion with a new bag
(3) Repeat: Repeat or re-program the infusion
VET-PRO VIP 2000™ Color Vision Operator Manual
to confirm the existing parameters.
to confirm the existing parameters.
to confirm the existing parameters and start
Intermittent
10:00
Interval Time
(HH:mm)
00:04
Adjust and press
Intermittent
10:00
KVO Rate
5
.
mL/hr
Adjust and press
Intermittent
10:00
Dose Rate
50 mL/hr
Dose Volume
3mL
Interval Time
0:04
Bag Volume
6 mL
Volume Left
6 mL
KVO Rate
5 mL/hr
Air Sensor
0.3 mL
Pressure
730 mmHg
Press
Key to Continue
Intermittent
10:00
Start Infusion ?
Press
key to Start
4 – Operation
80%
███■
-
Key
80%
███■
-
Key
80%
███■
-
80%
███■
-
56