CME VET-PRO VIP 2000 Operator's Manual page 38

Veterinary infusion pump system
Hide thumbs Also See for VET-PRO VIP 2000:
Table of Contents

Advertisement

Scroll to "More ..." and press
3.
4.
Enter Change Set up Level 1 code, using the numeric
keyboard and press
5.
Scroll to "Time Option" option. Press
Press
or
6.
7.
Press
twice.
8.
Volume Screen
Do one of the following:
Press
Enter the desired volume using the numeric keypad and
press
.
NOTE:
The screen will be blank upon initial setup.
9.
Infusion Time Screen
Enter the desired total time using the numeric keypad
(for example enter 0-2-0-0 for two hours) and press
.
NOTE:
First two digits represent hours and last two digits represent minutes.
VET-PRO VIP 2000™ Color Vision Operator Manual
.
.
, to turn Time Option ON.
to confirm the existing parameters.
Change Set up:
Pressure Level
Buzzer Level
More ...
Select
Technician Code
Press
Change Set up:
.
Air Bubble Size
Store Last Program
Keystroke beep
Time Option
WiFi module
Caregiver Delay
Language
Select
Continuous
Bag Volume
30
Adjust and press
Continuous
Total Time
02:00
Adjust and press
4 – Operation
,
and press
key
,
and press
80%
███■
10:00
-
mL
Key
80%
███■
10:00
-
(HH:mm)
Key
38

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents