System Overview
Technical Specifications
• Flow volume: 50 l/min
• Pressure: Maximum 8 bar (116 PSI)
Electrical Requirements
The line voltage must be free of power surges (overvoltage category II), power changes,
and loss of power for the protection of the electronic circuits.
Table 1.6
Electrical consumption
System current (SC)
SC and air compressor
Air compressor start
current
Voltage
Frequency
Power Line

Throughput and Configuration Specifications

Throughput
Sorting throughput
Aliquot throughput (ratio of
primary to secondary tubes
1:1, volume transfer 300 µl)
1-14
Electrical requirements
360 W (system fully loaded)
920 W (system fully loaded and air
compressor running)
1.5 A (system fully loaded)
4.0 A
10.0 A (~300 ms)
(approx. 10 times/hour if system is
fully loaded)
100-240 VAC [±10%]
50 / 60 Hz [±1%]
System and compressor, each 16 A
(Europe) or 20 A (North America)
AutoMate 1200 AutoMate 1250 AutoMate 2500 AutoMate 2550
800 tubes/h
AutoMate 1250
800 tubes/h
Up to 600
tubes/h
(primary and
secondary)
AutoMate 2550
480 W (system fully loaded)
1040 W (system fully loaded and air
compressor running)
2.2 A (system fully loaded)
4.7 A
10.0 A (~300 ms)
(approx. 10 times/hour if system is
fully loaded)
100-240 VAC [±10%]
50 / 60 Hz [±1%]
System and compressor, each 16 A
(Europe) or 20 A (North America)
1,200 tubes/h
1,200 tubes/h
Up to 600
tubes/h
(primary and
secondary)
B85841D