8 Appendix - Additional measuring sites | 8.3 Additional connection sites ACP X (additional earth connections)
1
Exemplary presentation of the measuring point on the PiezoLEDultrasonic scaler
① Test probe on ultrasonic scaler tip
Note
The switch on the handpiece must be activated during the EPA measure-
ment.
Note
Additional measuring points AP X need to be taken into consideration in the
presence of accessories: e.g. if third-party devices are connected, camera of
the multimedia system, etc.
8.3 Additional connection sites ACP X (additional earth
connections)
Note
A fixed connection from ACP to the protective earth conductor terminal (PE)
must be established for the EUL and EPA measurement. This can be accom-
plished, e.g., using a measuring cable and connection terminals.
in ultrasonic scaler handpiece
113 / 126
Instructions for use Primus 1058 Life