Display; Pressure Calibration Check - GE ApexPro TT Technical Manual

Designed to acquire and monitor physiological data for ambulating adult and pediatric patients within a defined coverage area.
Table of Contents

Advertisement

Display

Pressure calibration check

2001989-351A
Press and hold the
button while turning the switch on. This display includes
hardware and software version, the Gain Setting for the mic, and the pressure setting
(mmHg).
Gain Setting
The following test ensures the functionality of the Accutracker DX Noninvasive
Blood Pressure monitor.
1.
Connect the PDM200 or Mercury manometer to the patient cable/cuff
connection.
Sensym
PDM200M or
Equivalent
Manometer
Sample test setup
NOTE
Make sure no external pressure is applied to the Accutracker when you turn
it on.
2.
Press and hold the
Technical Calibrate mode and valve #1 is closed.
3.
Apply pressures to the Accutracker between 0 mmHg and 250 mmHg in
increments of 50 mmHg. Make sure there is no more than a +/–2 mmHg
difference between the mercury column or PDM200 display and the Accutracker
display value.
ApexPro™
Software Version
Hardware Version
Mercury Pressure in mm
Accutracker DX
Module to be
Tested
S U N T E C H
YES
NO
NEXT
LAST
INCR
DECR
A
ccutracker
D
S U N T E C H
R
START/
x
STOP
A B P M O N I T O R
Tubing,
Approximately
2 Feet
(pn 401582-001)
3-Way Tee
(pn 4745-101)
button and turn the Accutracker on. The unit is now in
Checkout
311A
Coupling, Male
(pn 400787-006)
Latex Bulb and
Deflation Valve
(by Baum)
(pn 0661-1890)
318A
7-21